Determination of the cross-contamination and validation of the cleaning process for an automated personalised dosing system
  1. Alexander Ferro Uriguen1,
  2. Idoia Beobide Telleria1,
  3. Silvia Martínez Arrechea1,
  4. Blanca Miró Isasi1,
  5. Carmen Sampedro Yangüela2,
  6. Mikel Urretavizcaya Anton3
  1. 1 Pharmacy Service, Hospital Ricardo Bermingham, San Sebastian, País Vasco, Spain
  2. 2 Alava Central Analysis Service, Lascaray Ikergunea Building, University of the Basque Country – Alava Campus, Vitoria Gasteiz, País Vasco, Spain
  3. 3 Pharmacy Service, Fundacion Onkologikoa, San Sebastian, Pais Vasco, Spain
  1. Correspondence to Alexander Ferro Uriguen, Pharmacy Service, Hospital Ricardo Bermingham, San Sebastian, País Vasco, Spain; alex.ferro{at}matiafundazioa.eus

Abstract

Objective To determine the degree of cross-contamination and to validate a cleaning process for an Automated Personalised Dosing System (APDS), respecting the permitted residue transfer limits.

Method An analytical determination by high pressure liquid chromatography with mass spectrometry was performed for 11 drugs considered critical for their potential for contamination, toxicity (National Institute for Occupational Safety and Health (NIOSH) List 2016) and/or narrow therapeutic index. The test was carried out for three consecutive weeks, seven repetitions per week, in a state of maximum contamination and maximum cleanliness. The main validation criterion of the cleaning protocol was to quantify that the active ingredients measured were below the Permitted Daily Exposure index or fraction 1/1000 of the minimum daily dose, both in a state of maximum contamination and when the APDS is in a clean condition, for an average number of tablets consumed per person per day.

Results In all the samples analysed, some compound could always be quantified below 2% of the permitted transfer limits, highlighting the presence of paracetamol and carbamazepine. In the state of maximum contamination, the mean±SD values of paracetamol and carbamazepine were 5.83±1.56 µg and 0.22±0.07 µg, respectively, without significant differences in mean values over the 3 weeks of the study. After the cleaning protocol was executed, only paracetamol and carbamazepine were detected, with the average values being 4.67 µg (95% CI 3.92 to 5.43, p<0.05) and 0.07 µg (95% CI 0.03 to 0.10, p<0.05) lower than the state of maximum contamination, respectively.

Conclusions In all cases the level of cross-contamination for an APDS was below the pre-established limits. The cleaning protocol has been validated, confirming APDS decontamination of the most critical medicines.

  • hospital pharmacy automation
  • risk management
  • dispensing robots
  • drug packaging
  • quality in health care
  • audit, self-inspection

Data availability statement

All data relevant to the study are included in the article or uploaded as supplementary information.

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Data availability statement

All data relevant to the study are included in the article or uploaded as supplementary information.

View Full Text