PKP-034 Determination of methotrexate in CSF by chemiluminescence using the architect
  1. P Montejano-Hervás,
  2. I Reyes-Torres,
  3. MD Aumente
  1. Hospital Universitario Reina Sofia, Farmacia Hospitalaria, Cordoba, Spain

Abstract

Background Intrathecal administration of methotrexate (MTX) in the treatment of different neoplastic diseases to prevent relapses in the CNS and the need to keep MTX concentrations in CSF during infusion of MTX requires study of the pharmacokinetics of MTX at this level. It is therefore necessary to measure concentrations of MTX in CSF.

Purpose To evaluate the validity (selectivity, accuracy and precision) of MTX measurements in CSF using a chemiluminescence assay for determination of MTX in plasma and to discard the matrix effect which might occur when measuring MTX in CSF.

Material and methods Different concentrations of MTX were tested in different types of samples to evaluate the selectivity and discard the matrix effect. To the studied matrix (CSF) were added increasing and known amounts of plasma with MTX (addition standard) and the results were correlated with the regression equation of the data obtained in both matrices. Accuracy was obtained by comparing the results of both matrices and obtaining the relative error. Due to the limitation of the small volume of CSF obtained in each extraction we were unable to perform different analyses on the same sample of CSF. 10 measurements on the same CSF sample containing a known quantity of MTX (obtained in our own test control) to evaluate precision were performed.

Results Factor analysis of standard additions gave the regression line X=0.00233+1.0105Y. The correlation coefficient was r = 0.99 and the relative error was -2.3%. The series of 10 repetitions of CSF with a known concentration (436 μM) gave an average value of 445.44 μM (424–464 μM), with SD of 12.5 and a variation coefficient of 2.8%.

Conclusion The good selectivity, accuracy and precision of the CFS analysis by CMIA (Architect) gave reliable data on concentrations of MTX in CSF, highlighting the absence of the matrix effect.