Table 1

Summary of the spectrophotometric data (660 nm) and visual appearance data for test Chemfort SAL/LL syringe combination and control blind hub/syringe combinations

Test Chemfort SAL/LL syringe combination resultsControl blind hub/LL syringe combination results
Test syringe sizeAverage absorbance at 660 nm (n=20)95% CIUnits tested (n)Spectrometer
pass/fail
Visual
pass/fail
Average absorbance at 660 nm95% CIUnits tested (n)Spectrometer
pass/fail
Visual pass/fail
50 mL−0.0040.00120PassPass3.3521.0945FailFail
20 mL−0.0040.00121PassPass1.9444.6823FailFail
1 mL−0.0050.00220PassPass0.048N/A1FailFail
  • LL, Luer-Lock; SAL, syringe adaptor lock.